At a Glance
- 6-Digit LED Display (±199999)
- Input: up to 40mV/V
- 20-bit A/D Converter
- Environmental Protection: IP65 (when installed)
- 10Vdc Load Cell Excitation @ 120mA max. Powers up to 4x 350Ω Load Cells
- Options: Alarm Relays, Analogue & Digital Outputs
- Customisation Of Software
- New Single Layer Menu for Easier Setup
- Compatible With Existing INT2 Series
- Improved Accuracy – 10 Point Linearisation
- Modular Construction Saves You Money
- Superior Resolution Guaranteed – 20-bit A/D Converter
- Dust Tight & Low-Pressure Water Jet Protection
- Waterproof Front Panel Cover Optional
- Perfect Replacement for Druck DPI 280 & 281 Indicators & INT2 Series
Description
Note: All DC-powered units are currently on a lead time of 4-5 weeks due to component shortages.
Applied Measurements intuitive4-L load cell digital indicator utilises a modular construction technique enabling it to be configured with the exact functionality required for your application. This flexible modular build ensures that you only pay for the features that you need. The INT4-L is the latest version of Applied Measurements INT2-L featuring a single layer menu structure for easier setup and improved load cell and process filtering.
The load cell readout is provided by a 6-digit LED display (also configurable with 3, 4 or 5 digits if preferred) coupled with an ultra-high-resolution 20-bit A/D converter and highly stable strain gauge bridge excitation source to provides superb resolution and stability in any application. The INT4-L has 10 point linearisation for outstanding accuracy.
Options available include voltage or current analogue outputs, 2 or 4 alarm relays and serial data output in one of several formats including RS232 ASCII, RS485 ASCII and RS485 Modbus RTU, making this a truly flexible load cell digital indicator.
The INT4-L is fully compatible with the INT2 series with its same front panel design, dimensions and connections, enabling an easy transition to the newer versions.
Should you require portability, our compact desktop enclosure the PCC, is also available to compliment the Intuitive4 load cell digital indicator or any other model from the Intuitive4 range. In addition, the SPC4 an IP67 immersion guaranteed front panel cover is available for harsh environments.
As its name implies, our Intuitive4-L load cell digital indicator is simple to install and calibrate using just the 4 front panel buttons and on-display prompts. The old ‘easy’ and ‘advanced’ menu systems have been made into one layer to make it simpler to use.
Have a look at our selection of load cells that can be used with this digital indicator.
Technical Specifications
CHARACTERISTICS | INTUITIVE4-L | UNITS |
---|---|---|
Display | 6 digit, 7 segment LED, 14mm high (-199999 to +199999) | |
Resolution | 1:400,000 max. (bipolar) | |
Display Update Rate: | 10/second | |
Bridge Excitation | 10.0±0.1% (option: 5.0±0.1%) | Volts dc |
Maximum Excitation Current | 120 | mA |
Input Range | 0.5-4 (contact sales for other options) | mV/V for max. display reading |
Input Impedance | >10 | Megaohms |
Digital Filtering | 0 to 5 (adjustable) | Seconds effective time constant |
Display Count By | 1, 2, 5, 10, 20, 50 digits (selectable) | |
Analogue Output (Optional) | ±10Vdc, 0-10Vdc or 4-20mA | |
Operating Temperature | 0 to +50 | ˚C |
Thermal Drift | Span = 25, Zero = 30 | ppm/˚C |
Operating Voltage | 100-240VAC (11-30VDC optional) | |
Environmental Protection | IP65 (IP67 optional with the SPC4 cover) | |
Weight | 300 | Grams |
Panel Cutout | 1/8 DIN (92mm x 45mm + I/-0) | |
Features | Tare facility, peak & valley detection, 10-point linearisation, 1 program memory (8 optional) | |
Trip Outputs (optional) | 2x or 4x relays rated 5A @ 250Vac, 2x SPCO, 2x or 4x solid state | |
Serial Data Output (optional) | RS232 (ASCII) / RS485 (ASCII or ModBus RTU) |
Product Dimensions
Model | Dimensions |
---|---|
PCC1 | W200mm x H65mm x D185mm |
PCC2 | W300mm x H65mm x D185mm |
PCC4 | W300mm x H145mm x D220mm |
PCC6 | W510mm x H145mm x D220mm |
PCC12 | W685mm x H240mm x D280mm |
Ordering Codes & Options
INT4-C-ANI-AL2-0-R-AC-MEM | INT4 | C | ANI | AL2 | 0 | R | AC | MEM |
---|---|---|---|---|---|---|---|---|
Product Family | ||||||||
INT4 | INT4 | |||||||
Input Version | ||||||||
L = Load Cell Input | L | |||||||
P = Process Input | P | |||||||
C = Counter/Rate Input | C | |||||||
Analogue Output | ||||||||
0 = None Fitted | 0 | |||||||
ANI = 4-20mA | ANI | |||||||
ANV = 0-10Vdc (uni-polar) | ANV | |||||||
ANB = ±10Vdc (bi-polar) | ANB | |||||||
Alarm Relays | ||||||||
0 = None Fitted | 0 | |||||||
AL2 = 2 x Alarm Relays | AL2 | |||||||
AL4 = 4 x Alarm Relays | AL4 | |||||||
SPCO = 2 x SPCO Relays | SPCO | |||||||
DSSAC = 2 x Solid State Relays (AC Version) - 100mA max @ 250Vac max | DSSAC | |||||||
DSSDC = 2 x Solid State Relays (DC Version) - 500mA max @ 60Vdc max | DSSDC | |||||||
QSSAC = 4x Solid State Relays (AC Version) - 100mA max @ 250Vac max | QSSAC | |||||||
QSSDC = 4x Solid State Relays (DC Version) - 500mA max @ 60Vdc max | QSSDC | |||||||
Serial Data Output | ||||||||
0 = None Fitted | 0 | |||||||
232 = RS232 ASCII | 232 | |||||||
485 = RS485 ASCII | 485 | |||||||
RTU = RS485 ModBus RTU | RTU | |||||||
Display Colour | ||||||||
R = Red | R | |||||||
RDLV = Red for Daylight Viewing | RDLV | |||||||
G = Green | G | |||||||
Supply Voltage | ||||||||
AC = 100-240Vac Universal Mains Supply | AC | |||||||
DC = 11-30Vdc | DC | |||||||
Additional & Special Features | ||||||||
0 = None | 0 | |||||||
100mV = 100mV Input (INT4-L only) | 100mV | |||||||
5VEXC = 5V Excitation Output | 5VEXC | |||||||
DIN = DIN Rail Mounting Option | DIN | |||||||
H = Real Time Clock Module | H | |||||||
HB = 1/2 Bridge Input for Load Cell Meters | HB | |||||||
MEM = 8 Calibration Memory (8 different load cells) - Requires 8-way Switch | MEM | |||||||
NAMUR = Namur Pulse Input for Counters/Ratemeters | NAMUR | |||||||
PBEXT = Connection for 4 External Programming Buttons | PBEXT | |||||||
PL = Plain Front Cover (no buttons) | PL | |||||||
POT = 3 Wire Potentiometer Input (INT4-P only) | POT | |||||||
SPC4 = IP67 Front Panel Cover | SPC4 | |||||||
xxxmV = Alternative Input Sensitivity on INT2-L (5, 10, 30, 100 or 300mV/V) | xxxmV | |||||||
Speak to Sales = Other special features will be determined by sales. | ???? |
Case Studies
Universal Fabric Tension Tester for All Fabrics and Styles
The Challenge - Accurately Measure Fabric Tension on Any Style of Furniture. Fabric strength testing guarantees the fabric being used on furniture meets the high quality needed for durability, comfort and support. Choosing the wrong type of fabric can lead to fabric tearing, loss of customer satisfaction and a decrease in sales and profit. Find out how our miniature s-beam load cell, intuitive4-L digital indicator and industrial LVDT displacement sensors were used to determine fabric strength.
S-Beam Load Cells Promise Accurate Pigment Dispensing
We chose Applied Measurements’ DBBSM s-beam load cells as the pigment is £1000 per tonne so has to be extremely accurate” says Paul Akers, Works Manager at PD Edenhall Ltd. One of the largest independent concrete facing brick manufacturer in the UK, needed a way to accurately monitor the amount of pigment being dispensed into the weigh hopper throughout the manufacture of concrete facing bricks. The pigment amount needs to be precise as inaccurate amounts of pigment can lead to incorrect colour blends leading to a loss in sales and profit. Read how our accurate DBBSM s-beam load cells and intuitive4-L digital indicators were used in this construction industry application.
How to Improve Aerodynamics with our Single Point Load Cells and Digital Indicators
The Scale Wind Tunnel uses two single point load cells and Intuitive2-L digital indicators enabling students to test the aerodynamics of scale models. Read how they did it.
Applied Measurements Builds Weighing System for Packaged Biomass Plantrooms
The weighing systems comprise of between 6 and 8 silo mount-style load cell weighing assemblies depending on the size of the plant, all of which are then wired into a bespoke junction box. The signal from the junction box feeds into an Intuitive2-L panel mounted digital display that provides a local weight reading in the biomass plantroom.